Aplicação
Medições da velocidade do ar e medições de vazão de ar.
Caracteristicas Técnicas
VELOCITY:
Range (100 mm head): 50 to 6,000 ft/min (0.25 to 30 m/s)
Accuracy (100 mm head): ±1.0% of reading or ±3 ft/min (±0.015 m/s),
whichever is greater
Range (50 mm head): 80 to 3,000 ft/min (0.4 to 15 m/s)
Accuracy (50 mm head): ±1.5% of reading or ±5 ft/min (±0.025 m/s),
whichever is greater
TEMPERATURE SENSOR:
Range: 32 to 176°F (0 to 80°C)
Resolution: 0.1°F (0.1°C)
Accuracy: ±1.8°F (±1.0°C)
INSTRUMENT TEMPERATURE RANGE:
Operating (electronics): 40 to 113°F (5 to 45°C)
Operating (vane head): 32 to 176°F (0 to 80°C)
Storage: -4 to 140°F (-20 to 60°C)
VOLUMETRIC FLOWRATE:
Range1: 0 to 1,230,000 l/s
0 to 4,360,000 m3/hr
0 to 2,604,000 ft3/min
DUCT SIZE:
Range: 1 to 250 inches in increments of 0.1 inches
(1 to 635 cm in increments of 0.1 cm)