• FHM-10

    Aplicação

    + Research Laboratories
    + Life Science and Pharmaceutical
    + Universities and Academic
    + Vivariums
    + Healthcare Facilities

    Caracteristicas Técnicas

    -Display Range
    0 to 1,000 fpm (0 to 5.08 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
    -Low Alarm Range
    5 to 960 fpm (0.03 to 4.88 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
    -High Alarm Range
    80 to 1,000 fpm (0.41 to 5.08 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
    -Analog Outputs
    0-10 VDC or 4-20 mA
    Represents Face Velocity, or Flow Rate
    -Alarm Contact Outputs
    SPST, 2A @ 30 VDC Nominal
    -Contact Inputs
    Night Setback, Flow

    FHM-10